VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Pro...
FDA Device Recall #Z-0834-2013 — Class II — September 28, 2012
Recall Summary
| Recall Number | Z-0834-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 28, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Ventana Medical Systems Inc |
| Location | Oro Valley, AZ |
| Product Type | Devices |
| Quantity | 54 instruments |
Product Description
VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.
Reason for Recall
Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu
Distribution Pattern
Worldwide distribution: USA (nationwide) and country of: Germany.
Lot / Code Information
Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).
Other Recalls from Ventana Medical Systems Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1374-2022 | Class II | Ventana HE 600 System, automated slide preparer... | Jun 1, 2022 |
| Z-0894-2022 | Class II | BenchMark ULTRA and DISCOVERY ULTRA Instruments | Jan 20, 2022 |
| Z-2262-2020 | Class II | Model Number: 06917259001 UDI: 0401563097601... | Oct 4, 2019 |
| Z-3014-2018 | Class I | INFORM HPV III Fam 16 Probe - US Export, Catalo... | Aug 2, 2018 |
| Z-2996-2018 | Class I | OptiView DAB IHC Detection Kit, Catalog Number ... | Aug 2, 2018 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.