Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part num...
FDA Device Recall #Z-1281-2015 — Class II — February 17, 2015
Recall Summary
| Recall Number | Z-1281-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 17, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Synthes, Inc. |
| Location | West Chester, PA |
| Product Type | Devices |
| Quantity | 183 |
Product Description
Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052). The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX. The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.
Reason for Recall
For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
Distribution Pattern
Nationwide Distribution
Lot / Code Information
Part numbers:03.010.048 and 03.010.052 with Lot Numbers: 4874015 4874021 4927262 4954481 4963192 4874016 4874022 4944750 4957586 4963193 4784017 4899564 4944751 4957587 4972003 4874018 4899565 4954478 4957588 4972004 4874019 4927260 4954479 4957589 4874020 4927261 4954480 4963190 4922533 4922975 4981158 4922970 4923015 4981159 4922971 4977241 4981160 4922972 4977242 4981162 4922973 4981156 5035905 4922974 4981157 5035906
Other Recalls from Synthes, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2698-2020 | Class II | MatrixNEURO Screws - Product Usage: The intende... | Jun 18, 2020 |
| Z-0880-2020 | Class II | 2.4MM TI VA LOCKING SCREW STARDRIVE 12MM STERIL... | Dec 16, 2019 |
| Z-0883-2020 | Class II | 2.4MM TI VA LOCKING SCREW STARDRIVE 24MM STERIL... | Dec 16, 2019 |
| Z-0881-2020 | Class II | 2.4MM TI VA LOCKING SCREW STARDRIVE 16MM STERIL... | Dec 16, 2019 |
| Z-0879-2020 | Class II | 2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERIL... | Dec 16, 2019 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.