Rheumatoid Factor Standard (RF CAL) Catalogue Number: RF2301 GTIN: 05055273205039 Product Usa...
FDA Device Recall #Z-0282-2020 — Class II — August 9, 2019
Recall Summary
| Recall Number | Z-0282-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 9, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Randox Laboratories Ltd. |
| Location | Crumlin (North), N/A |
| Product Type | Devices |
| Quantity | 17 kits |
Product Description
Rheumatoid Factor Standard (RF CAL) Catalogue Number: RF2301 GTIN: 05055273205039 Product Usage: The Randox Rheumatoid Factor Latex-Enhanced Immunorutbidimetric Test Kit is an in vitro diagnostic reagent for the quantitative determination of rheumatoid factors (RF) in human serum. This test kit may be used to facilitate the detection and diagnosis of Rheumatoid Arthritis. RF have also been observed in the serum of patients with systemic lupus eruthematosus, hepatitis. Liver cirrhosis and syphilis. However, in these conditions the RF titre is much lower than in rheumatoid arthritis. This test kit must be used by suitably qualified personnel under appropriate laboratory conditions. RF calibrator and RF positive control are for use in the calibration and quality control of RF latex-enhanced immunoturbidimetric assays.
Reason for Recall
The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
Distribution Pattern
US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
Lot / Code Information
Batch/Lot # 463479/1325 -1329RF 466715/1325 -1329RF 474059/1339 1343RF 480410/ 1347 1351RF
Other Recalls from Randox Laboratories Ltd.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2425-2024 | Class II | RX Series CYSC- Cystatin C Reagent -Intended fo... | Jun 7, 2024 |
| Z-1786-2024 | Class III | Evidence MultiSTAT with software v 3.7-Analyzer... | Apr 5, 2024 |
| Z-1759-2024 | Class II | Clinical Chemistry Calibration Serum (Catalog n... | Mar 5, 2024 |
| Z-1328-2024 | Class II | Liquid Protein Calibrators. C3 and Haptoglobin ... | Mar 1, 2024 |
| Z-1369-2024 | Class II | Cystatin C Calibrator - In vitro diagnostic (IV... | Feb 14, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.