Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, M...
FDA Device Recall #Z-1218-2019 — Class II — August 22, 2018
Recall Summary
| Recall Number | Z-1218-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 22, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Smiths Medical ASD Inc. |
| Location | Minneapolis, MN |
| Product Type | Devices |
| Quantity | 136,600 units |
Product Description
Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.
Reason for Recall
Several lots of Jelco¿ ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.
Lot / Code Information
Lot Numbers: 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512
Other Recalls from Smiths Medical ASD Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0147-2025 | Class II | smiths medical portex, Thermovent 1200, 15mm/22... | Jul 18, 2024 |
| Z-0145-2025 | Class II | smiths medical portex, Y' PIECE 15MM, REF 100/2... | Jul 18, 2024 |
| Z-0146-2025 | Class II | smiths medical portex, 'ORATOR' SPEAKING VALVE ... | Jul 18, 2024 |
| Z-0143-2025 | Class II | smiths medical portex, Nasopharyngeal Airway, S... | Jul 18, 2024 |
| Z-0144-2025 | Class II | smiths medical portex, 15mm Double Swivel Conne... | Jul 18, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.