J-Plasma Precise(R) OPEN, Catalog Numbers: BVX-044-BPP, BVX-044-BPS, BVX-150-BPP, BVX-150-BPS
FDA Device Recall #Z-1153-2018 — Class II — February 8, 2018
Recall Summary
| Recall Number | Z-1153-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 8, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Bovie Medical Corporation |
| Location | Clearwater, FL |
| Product Type | Devices |
| Quantity | 1005 units |
Product Description
J-Plasma Precise(R) OPEN, Catalog Numbers: BVX-044-BPP, BVX-044-BPS, BVX-150-BPP, BVX-150-BPS
Reason for Recall
Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.
Distribution Pattern
worldwide
Lot / Code Information
All lots manufactured since 01/01/2014
Other Recalls from Bovie Medical Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1154-2021 | Class II | HNM Disposable Bipolar Ablators - Product Usage... | Feb 1, 2021 |
| Z-1153-2021 | Class II | Velocity Disposable Bipolar Ablators - Product ... | Feb 1, 2021 |
| Z-1155-2021 | Class II | ConMed Disposable Bipolar Ablators - Product Us... | Feb 1, 2021 |
| Z-2061-2020 | Class II | Renuvion/ J-Plasma Precise Handpiece. Catalog ... | Apr 20, 2020 |
| Z-2060-2020 | Class II | Renuvion/ J-Plasma Precise Open Handpiece. Cat... | Apr 20, 2020 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.