Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF2...

FDA Recall #Z-1293-2018 — Class II — December 15, 2017

Recall #Z-1293-2018 Date: December 15, 2017 Classification: Class II Status: Terminated

Product Description

Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.

Reason for Recall

if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Recalling Firm

GE Healthcare, LLC — Waukesha, WI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

355

Distribution

Worldwide - US Nationwide Distribution.

Code Information

Model # - 105-PT-001 ( all units distributed from October 2011 - May 2016)

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.