MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF...

FDA Device Recall #Z-1603-2018 — Class II — March 26, 2018

Recall Summary

Recall Number Z-1603-2018
Classification Class II — Moderate risk
Date Initiated March 26, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Philips Medical Systems Nederlands
Location Best, N/A
Product Type Devices
Quantity 184

Product Description

MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Reason for Recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Lot / Code Information

System Serial # 32 177 259 239 264 258 52 67 111 257 94 142 121 191 143 260 265 74 207 101 182 174 124 263 4 172 254 151 246 64 183 76 152 188 261 242 102 34 147 241 134 178 245 176 175 38 162 107 220 65 194 69 247 13 215 78 112 29 139 98 79 158 61 33 106 35 50 16 80 205 161 11 204 37 12 189 9 232 15 228 155 47 171 213 209 218 96 90 212 217 181 163 140 114 115 108 71 141 14 36 48 224 25 105 57 216 138 24 28 211 160 192 43 44 157 154 41 219 8 68 129 75 221 7 10 63 146 45 26 125 193 210 1 60 3 132 81 22 23 208 5 131 135 88 226 6 156 49 91 92 93 201 250 251 72 95 148 39 31 2 30 51 100 255 252 256 109 196 164 170 227 173 83 159 53 137 54 116 118 119 167 168 169 97

Other Recalls from Philips Medical Systems Nederlands

Recall # Classification Product Date
Z-1240-2019 Class II Ingenia Ambition S (781359), Ingenia Ambition X... Mar 25, 2019
Z-1141-2019 Class II Foot Switches used with the following systems: ... Feb 19, 2019
Z-1140-2019 Class II Foot Switches used with Philips MultiDiagnost E... Feb 19, 2019
Z-0903-2019 Class II SmartPath to dStream for XR and 3.0T, Model Num... Dec 21, 2018
Z-0908-2019 Class II Intera 1.5T High Performance, Model Number 7812... Dec 21, 2018

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.