Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
FDA Recall #Z-1663-2025 — Class II — March 27, 2025
Product Description
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Reason for Recall
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Recalling Firm
IHB OPERATIONS B.V. — Zwolle, N/A
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
112 units
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Code Information
UDI/DI 10841736116071, All Serial Numbers
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.