Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN ...

FDA Recall #Z-0707-2020 — Class II — December 6, 2018

Recall #Z-0707-2020 Date: December 6, 2018 Classification: Class II Status: Terminated

Product Description

Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan). The percutaneous pin adapter is a component of the StealthAir Frame Assembly.

Reason for Recall

Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.

Recalling Firm

Medtronic Navigation, Inc. — Louisville, CO

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

1,297 percutaneous pin adapters

Distribution

Distribution was nationwide. There was government/military/foreign distribution.

Code Information

Percutaneous pin adapter lot numbers: 150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.