Discovery RT labeled as: a. MID BJG; b. HVY BJG
FDA Device Recall #Z-2581-2018 — Class II — May 24, 2018
Recall Summary
| Recall Number | Z-2581-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 24, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | GE Healthcare, LLC |
| Location | Waukesha, WI |
| Product Type | Devices |
| Quantity | 42 units total |
Product Description
Discovery RT labeled as: a. MID BJG; b. HVY BJG
Reason for Recall
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Distribution Pattern
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Lot / Code Information
Model Number 2374682-17. a. MID BJG. Serial Number CBCIG1800017HM (System ID 847695CTRT, UDI 01008406821186991118020021CBCIG1800017HM), Serial Number CBCIG1700070HM (System ID 251665MCIRT, UDI 01008406821186991117120021CBCIG1700070HM), Serial Number CBCIG1700064HM (System ID 973831RT, UDI 01008406821186991117120021CBCIG1700064HM), Serial Number CBCIG1800001HM (System ID 708747RT590, UDI 01008406821186991118010021CBCIG1800001HM); b. HVY BJG. Serial Number CBCIG1700062HM (System ID 904819RTCT, UDI 01008406821186991117120021CBCIG1700062HM), Serial Number CBCIG1700040HM (System ID 843692GSOCT, UDI 01008406821186991117100021CBCIG1700040HM)
Other Recalls from GE Healthcare, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2022 | Class II | (1) GE Centricity Universal Viewer 6.0, Model 2... | Jul 15, 2022 |
| Z-1430-2022 | Class II | SIGNA Premier magnetic resonance scanner, model... | Jun 24, 2022 |
| Z-1466-2022 | Class II | Centricity PACS Software Version 7.0 SP0.0.4.7 | Jun 14, 2022 |
| Z-1333-2022 | Class II | GE Centricity Universal Viewer Zero Footprint. ... | Jun 8, 2022 |
| Z-1511-2022 | Class II | Service manuals for the CT, PET, or NM table as... | Jun 2, 2022 |
Frequently Asked Questions
Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.