Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used...
FDA Device Recall #Z-1709-2019 — Class II — April 15, 2019
Recall Summary
| Recall Number | Z-1709-2019 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 15, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Abbott Gmbh & Co. KG |
| Location | Wiesbaden |
| Product Type | Devices |
| Quantity | 19,184 |
Product Description
Alinity c Cuvette Segments, List Number 04S47-01 Product Usage: The Alinity c Cuvettes are used on the Alinity c system processing module. The Alinity c processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology.
Reason for Recall
Individual cuvettes within the Alinity c Cuvette Segment may become seated lower than the designed height. This may result in inadequate dispense into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.
Distribution Pattern
US Nationwide Distribution- AL, AR,CA, FL, GA, IL, LA, MA, ME, MN, MO, NJ, NY, OK, SC, TN, TX, UT, & VA WORLDWIDE - Albania, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Columbia, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, Norway, Pakistan, Philippines, Poland, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Turkey, United Kingdom, and Vietnam.
Lot / Code Information
All serial numbers
Other Recalls from Abbott Gmbh & Co. KG
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2701-2020 | Class II | Alinity ci series System Control Module (SCM); ... | Jun 19, 2020 |
| Z-2401-2020 | Class II | Alinity C, Processing Module. Chemistry analyz... | May 12, 2020 |
| Z-2117-2020 | Class II | ARCHITECT iGentamicin Reagent Kit - Product Usa... | Apr 24, 2020 |
| Z-1728-2020 | Class II | ARCHITECT HAVAB G Reagent Kit. List Number 6L27... | Mar 9, 2020 |
| Z-1580-2020 | Class II | LN 3R70-01; software version 2.6.2 and earlier.... | Jan 24, 2020 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.