Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting ...

FDA Device Recall #Z-1350-2023 — Class II — January 25, 2023

Recall Summary

Recall Number Z-1350-2023
Classification Class II — Moderate risk
Date Initiated January 25, 2023
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Church & Dwight Inc
Location Princeton, NJ
Product Type Devices
Quantity 780000 downloads

Product Description

Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)

Reason for Recall

Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.

Distribution Pattern

US Nationwide distribution.

Lot / Code Information

All lots with the Easy Read App

Other Recalls from Church & Dwight Inc

Recall # Classification Product Date
Z-1811-2013 Class II Sodium Bicarbonate, USP Provided bulk as raw ... Jun 11, 2013
Z-2093-2012 Class II Arm and Hammer Spinbrush Pro Clean Sonic Powere... Dec 22, 2011
Z-2094-2012 Class II Arm and Hammer Spinbrush Pro Clean Sonic Color ... Dec 22, 2011
Z-2095-2012 Class II Arm and Hammer Spinbrush Pro White Sonic Crest ... Dec 22, 2011
Z-0858-2012 Class II Arm and Hammer Spinbrush Dual Action Powered To... Dec 22, 2011

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.