Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body ...

FDA Device Recall #Z-0322-2017 — Class II — September 1, 2016

Recall Summary

Recall Number Z-0322-2017
Classification Class II — Moderate risk
Date Initiated September 1, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic Navigation, Inc.
Location Louisville, CO
Product Type Devices
Quantity 100 total units

Product Description

Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures

Reason for Recall

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Distribution Pattern

Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.

Lot / Code Information

Lot number 151481

Other Recalls from Medtronic Navigation, Inc.

Recall # Classification Product Date
Z-1022-2025 Class II Medtronic PIN, 100mm STERILE PERCUTANEOUS REF: ... Dec 4, 2024
Z-0050-2025 Class II StealthStation S8 App versions 1.3.0 and 1.3.2,... Aug 22, 2024
Z-1776-2024 Class I Stealth S8 Clinical Software Application, REF: ... Mar 26, 2024
Z-0698-2024 Class II Medtronic, REF 9733235, Pin, 100mm, STERILE, PE... Nov 28, 2023
Z-0699-2024 Class II Medtronic, REF 9733236, Pin, 150mm, STERILE, PE... Nov 28, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.