FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
FDA Recall #Z-0628-2016 — Class III — December 14, 2015
Product Description
FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
Reason for Recall
Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
Recalling Firm
Becton, Dickinson and Company, BD Biosciences — San Jose, CA
Classification
Class III — Not likely to cause adverse health consequences.
Product Type
Devices
Product Quantity
1,388 units
Distribution
Nationwide Distribution
Code Information
Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.