Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for ext...
FDA Device Recall #Z-0004-2017 — Class I — September 1, 2016
Recall Summary
| Recall Number | Z-0004-2017 |
| Classification | Class I — Serious risk |
| Date Initiated | September 1, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Leonhard Lang Medizintechnik GmbH |
| Location | Innsbruck, N/A |
| Product Type | Devices |
| Quantity | 11,110 ( US 8,040; OUS 3,070) |
Product Description
Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.
Reason for Recall
There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Distribution Pattern
Worldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.
Lot / Code Information
US: 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 50130-0777; 41023-0771; 41008-0778; 40730-0778; 40618-0778; and, 40130-0776. Outside US: 60725-0774; 60620-0776; 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 40827-0777; 40730-0778; 40618-0778; and, 40130-0776.
Other Recalls from Leonhard Lang Medizintechnik GmbH
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-3109-2017 | Class I | Skintact Electrode for Defibrillation, Pacing, ... | Aug 8, 2017 |
| Z-0292-2016 | Class II | ECG Electrodes: 50007 Stablebase SBW55 50... | Nov 2, 2015 |
| Z-1350-2015 | Class II | Skintact ECG Electrode FS-50, Catalog no. 5802... | Feb 24, 2015 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.