Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra...
FDA Recall #Z-0604-2016 — Class II — December 18, 2015
Product Description
Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
Reason for Recall
When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
Recalling Firm
Elekta, Inc. — Atlanta, GA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
154
Distribution
Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.
Code Information
Versions 3.3 and higher
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated