Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra...

FDA Device Recall #Z-0604-2016 — Class II — December 18, 2015

Recall Summary

Recall Number Z-0604-2016
Classification Class II — Moderate risk
Date Initiated December 18, 2015
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Elekta, Inc.
Location Atlanta, GA
Product Type Devices
Quantity 154

Product Description

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Reason for Recall

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.

Lot / Code Information

Versions 3.3 and higher

Other Recalls from Elekta, Inc.

Recall # Classification Product Date
Z-2006-2026 Class II Brand Name: Leksell Vantage Arc System REF: 10... Mar 19, 2026
Z-1581-2026 Class II Brand Name: Leksell Gamma Knife Product Descri... Feb 18, 2026
Z-1164-2026 Class II Brand Name: Elekta Medical Linear Accelerator ... Dec 14, 2025
Z-1339-2026 Class II Brand Name: Leksell GammaPlan Product Name: Ra... Dec 12, 2025
Z-1340-2026 Class II Brand Name: Leksell GammaPlan Product Name: Ra... Dec 12, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.