Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra...

FDA Recall #Z-0604-2016 — Class II — December 18, 2015

Recall #Z-0604-2016 Date: December 18, 2015 Classification: Class II Status: Terminated

Product Description

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Reason for Recall

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

Recalling Firm

Elekta, Inc. — Atlanta, GA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

154

Distribution

Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.

Code Information

Versions 3.3 and higher

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated