Perkin Elmer ClWIZARD2 5-detector, 550 samples Product Code: 2470-0050, 3470-0050. The Wizard2 ...

FDA Recall #Z-0634-2016 — Class II — December 8, 2015

Recall #Z-0634-2016 Date: December 8, 2015 Classification: Class II Status: Terminated

Product Description

Perkin Elmer ClWIZARD2 5-detector, 550 samples Product Code: 2470-0050, 3470-0050. The Wizard2 gamma counter is intended to detect and count gamma radiation emitted by clinical samples.

Reason for Recall

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

Recalling Firm

Perkinelmer — Waltham, MA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

62 units

Distribution

Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.

Code Information

Model: 2470-0050 Serial Numbers: DG05106329 DG02117493 DG11107166 DG09106782 DG07118133 DG07118141 DG09106791 DG04117727 DG06118036 DG12107192 DG02106101 DG04106223 DG01128843 DG12107272 DG07118124 DG09106892 DG12118684 DG12118673 DG11107126 DG04117704 DG04117763 DG01117363 DG06106484 DG12107240 DG01128871 DG05117880 DG12118685 DG09118357 DG03106204 DG05106386 DG11107137 DG03106172 DG03106206 DG02106113 DG09106818 DG02106065 DG04106309 DG09106882 DG11118590 DG06106483 DG01117316 DG08118221 DG06117898 DG12107263 DG08118250 DG12118711 DG04117710 DG06106448 DG06106453 DG06106463 DG06117912 DG03106135 DG01117373 DG05117837 DG06106517 DG01128805 DG06117970 DG04106310 DG02117521 DG06106538 Model: 3470-0050 (2 units) Serial Numbers: DG12118660 DG10118464

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.