Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
FDA Device Recall #Z-0597-2016 — Class II — November 18, 2015
Recall Summary
| Recall Number | Z-0597-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 18, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Siemens Medical Solutions USA, Inc |
| Location | Malvern, PA |
| Product Type | Devices |
| Quantity | 73 units |
Product Description
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Reason for Recall
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Distribution Pattern
Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
Lot / Code Information
Software Version: SW VD-10E
Other Recalls from Siemens Medical Solutions USA, Inc
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1175-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1174-2026 | Class II | Interventional Fluoroscopic X-Ray System: LUMIN... | Dec 29, 2025 |
| Z-1241-2026 | Class II | NAEOTOM Alpha.Prime Software applications: s... | Dec 19, 2025 |
| Z-1240-2026 | Class II | NAEOTOM Alpha Software applications: syngo.C... | Dec 19, 2025 |
| Z-1239-2026 | Class II | SOMATOM X.ceed Software applications: syngo.... | Dec 19, 2025 |
Frequently Asked Questions
A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.