Perkin Elmer WIZARD2 5-detector, 1000 samples Product Code: 2470-0150, 3470-0150 The Wizard gam...
FDA Recall #Z-0636-2016 — Class II — December 8, 2015
Product Description
Perkin Elmer WIZARD2 5-detector, 1000 samples Product Code: 2470-0150, 3470-0150 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
Reason for Recall
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
Recalling Firm
Perkinelmer — Waltham, MA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
4 units
Distribution
Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.
Code Information
Serial Numbers: DG04129084 DG12118776 DG04117773 DG05106408
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated