Product 33 consists of all product under product code: JDI and same usage: Item no: 662406515...

FDA Recall #Z-1131-2016 — Class II — January 11, 2016

Recall #Z-1131-2016 Date: January 11, 2016 Classification: Class II Status: Terminated

Product Description

Product 33 consists of all product under product code: JDI and same usage: Item no: 662406515 HGP II ACETABULAR CUP BON 662406520 HGP II ACETABULAR CUP BON 662406525 HGP II ACETABULAR CUP BON 662406530 HGP II ACETABULAR CUP BON 662406535 HGP II ACETABULAR CUP BON 662406540 HGP II ACETABULAR CUP BON 662406550 HGP II ACETABULAR CUP BON 662406560 HGP II ACETABULAR CUP BON For use in total hip arthroplasty

Reason for Recall

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Recalling Firm

Zimmer Manufacturing B.V. — Mercedita, PR

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

2236

Distribution

US Nationwide Distribution

Code Information

lot no.: 61674472 61906696 62016799 62332483 61674473 61902194 61906697 62326386 61674476 61953546 62161131 62172656 62326385 61953547 62011539 62017894 62130757 61886072 61953548 62017895 62172659 61886073 61948763 61953551 62172661 61874793 61948764 61996232 62017897 62172663 61874794 61953554 62286122

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated