Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised o...
FDA Device Recall #Z-0799-2016 — Class I — January 6, 2016
Recall Summary
| Recall Number | Z-0799-2016 |
| Classification | Class I — Serious risk |
| Date Initiated | January 6, 2016 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Cook Inc. |
| Location | Bloomington, IN |
| Product Type | Devices |
| Quantity | 1341 |
Product Description
Single Lumen Central Venous Catheter Sets and Trays The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.
Reason for Recall
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.
Distribution Pattern
Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada
Lot / Code Information
RPN/Catalog number: C-PMS-250-PED C-PMS-251J-PED C-PMS-300-CHILDRENS-032285 C-PMS-301J-PED C-PMS-301J-PED-BH C-PMS-301-PED C-PMSY-250 C-PMSY-251 C-PMSY-251J C-PMSY-300 C-PMSY-300J C-PMSY-300J-UT-A C-PMSY-300-PED C-PMSY-301J C-PMSY-301J-BH C-PMSY-301J-PED C-PMSY-301J-PED-OTP C-PMSY-400-HURLEY-022388 C-PUM-300J C-PUM-301 C-PUM-301J C-PUMY-301J Lots: NS6168054 NS6178118 NS6063352 NS5855441 5824486 5898058 5967405 6079119 NS5932737 NS6028554 NS6042799 NS5922707 NS5932811 NS6145002 5987014 NS6178099 NS6028540 NS5830758 NS5866960 NS5918174 NS5974263 NS6042577 NS6047450 NS6090278 NS6105922 NS6124990 NS5881337 NS5892116 NS6031991 5782006 5855241 NS5881336 NS5892113 NS5945282 6062729 NS5970950 NS6108255 NS6153173 5793198 5925744 6010406 6148800 NS5796483 NS5928602 NS6047451 NS5970942 NS5954223 5945120 NS5954224 5973354
Other Recalls from Cook Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0388-2022 | Class II | Flexor Check-Flo Introducer Ansel Modification ... | Nov 12, 2021 |
| Z-0387-2022 | Class II | Flexor Check-Flo Introducer Ansel Modificatio... | Nov 12, 2021 |
| Z-0414-2022 | Class II | Skinny Needle with Chiba Tip. RPN 090010, 09001... | Oct 13, 2021 |
| Z-0222-2022 | Class II | Transseptal Needle, Trocar. RPN TSNC-18-71.0, T... | Oct 8, 2021 |
| Z-0223-2022 | Class II | Transseptal Needle with Catheter. RPN TSN-17-75... | Oct 8, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.