Ergon 2 Articulating Equipment Boom Monitor Brackets mounting point for display interfaces uti...
FDA Recall #Z-0655-2016 — Class II — December 2, 2015
Product Description
Ergon 2 Articulating Equipment Boom Monitor Brackets mounting point for display interfaces utilized in clinical applications for viewing purposes and to provide an articulating point of movement on the ceiling mounted radial arm suspension system of the Ergon 2 equipment pendant / booms
Reason for Recall
The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket. A display interface falling from its mounting location may result in injury to the patient or provider.
Recalling Firm
Skytron, Div. The KMW Group, Inc — Grand Rapids, MI
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
286
Distribution
Worldwide Distribution - US including WI, TX, FL, NC, MD, OH, AZ, CA, UT, MT, OK, HI, CO, WY, MO, NJ, IL, NY, MA, IA, DE, PA, MI, IN, and WV and Internationally to Canada, and Japan.
Code Information
Model Number(s): H8-010-50-FC2-LH and H8-010-50-FC2-RH. VESA plates distributed from 09/26/08 to 04/30/13.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.