Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOM...
FDA Recall #Z-0486-2026 — Class II — September 25, 2025
Product Description
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
Reason for Recall
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Recalling Firm
Covidien, LP — North Haven, CT
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
2 units (OUS only)
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Code Information
CFN: BOX06170V4; UDI-DI: 00763000988975; Lot numbers: 0231961695;
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated