Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix ...

FDA Recall #Z-2596-2023 — Class II — August 7, 2023

Recall #Z-2596-2023 Date: August 7, 2023 Classification: Class II Status: Ongoing

Product Description

Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.

Reason for Recall

Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

Recalling Firm

Stryker Corporation — Portage, MI

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

109 units US; 290 units OUS

Distribution

Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK

Code Information

GTIN: 04546540039415 Lot Numbers: 22314012 22336012

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.