iChem¿VELOCITY Urine Chemistry System-

FDA Device Recall #Z-1944-2020 — Class II — March 30, 2020

Recall Summary

Recall Number Z-1944-2020
Classification Class II — Moderate risk
Date Initiated March 30, 2020
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Beckman Coulter Inc.
Location Brea, CA
Product Type Devices
Quantity 1557 units

Product Description

iChem¿VELOCITY Urine Chemistry System-

Reason for Recall

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

Distribution Pattern

US Distribution: Nationwide Foreign Distribution: Worldwide

Lot / Code Information

Catalog #700-7176-001 and #700-7177-001, All part numbers

Other Recalls from Beckman Coulter Inc.

Recall # Classification Product Date
Z-0541-2026 Class II DxFLEX Flow Cytometer Catalog Number C78500 UD... Sep 22, 2025
Z-0441-2026 Class II SYNCHRON Systems Direct LDL Cholesterol (LDLD) ... Sep 17, 2025
Z-0071-2026 Class II SYNCHRON Systems Phosphorus (PHOSm) Reagent RE... Sep 3, 2025
Z-2573-2025 Class II UniCel DxH 690T COULTER Cellular Analysis Syste... Aug 13, 2025
Z-2574-2025 Class II UniCel DxH 800 COULTER Cellular Analysis System... Aug 13, 2025

Frequently Asked Questions

Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.