MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled ...
FDA Device Recall #Z-1944-2012 — Class II — June 8, 2012
Recall Summary
| Recall Number | Z-1944-2012 |
| Classification | Class II — Moderate risk |
| Date Initiated | June 8, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Impac Medical Systems Inc |
| Location | Sunnyvale, CA |
| Product Type | Devices |
| Quantity | 11 affected sites |
Product Description
MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086
Reason for Recall
Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.
Distribution Pattern
Worldwide Distribution -- USA, including the states of VA, CT, GA, NJ and NY and the countries of Belgium, Denmark, France, and the United Kingdom.
Lot / Code Information
Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7.
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.