Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in ...
FDA Device Recall #Z-2053-2020 — Class II — April 13, 2020
Recall Summary
| Recall Number | Z-2053-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 13, 2020 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Philips North America, LLC |
| Location | Andover, MA |
| Product Type | Devices |
| Quantity | 94,034 units |
Product Description
Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.
Reason for Recall
The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.
Distribution Pattern
Worldwide distribution - US Nationwide distribution.
Lot / Code Information
All units. Product has been discontinued.
Other Recalls from Philips North America, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0926-2022 | Class II | The IFU for the Autoclavable Temperature Probes... | Feb 7, 2022 |
| Z-1326-2021 | Class II | Philips SureSigns VM4, VM6 and VM8 - Product Us... | Mar 23, 2021 |
| Z-0852-2021 | Class II | Emission Computed Tomography System Image Proce... | Nov 19, 2020 |
| Z-0285-2021 | Class II | Philips Sterilizable Defibrillator Paddles, Swi... | Sep 14, 2020 |
| Z-0284-2021 | Class II | Philips Sterilizable Defibrillator Paddles, Swi... | Sep 14, 2020 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.