Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiol...
FDA Recall #Z-0801-2014 — Class II — December 4, 2013
Product Description
Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Philips Healthcare System, Cleveland, OH. Produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for Recall
Machine Name in TumorLOC Basic Mode is obscured. When TumorLOC is in Basic Mode and Set Vertical Laser equals Yes, the Machine Name label and field are covered by the controls above it. The Machine Name cannot be viewed or changed.
Recalling Firm
Philips Medical Systems (Cleveland) Inc — Cleveland, OH
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
32 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, MD, NH, NV, OH, and OR, and the countries of Australia, China, France, India, Indonesia, Italy, Korea, Mexico, Poland, Romania, Russia, Spain, Switzerland, Taiwan, Thailand & United Arab Emirates.
Code Information
Model #: 728243; 7133, 7170, 7225, 7262, 7265, 7701, 7706, 7779, 7788, 7792, 7797, 7802, 7805, 7806, 7807, 7808, 7810, 7811, 7812, 7814, 7815, 7816, 7824, 7831, 7832, 7833, 7834 ,7835. Model #728244; 7621, 7636, 7809 & 7844.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated