RTH8 Rotor, P/N X01-005847-001. RTH8 used in the StatSpin Express 4 Horizontal Centrifuge, Model...

FDA Recall #Z-0857-2015 — Class II — December 3, 2014

Recall #Z-0857-2015 Date: December 3, 2014 Classification: Class II Status: Terminated

Product Description

RTH8 Rotor, P/N X01-005847-001. RTH8 used in the StatSpin Express 4 Horizontal Centrifuge, Model M510, Product No. SSH4. The RTH8 rotor is used with the Statspin Express 4 Horizontal Centrifuge. StatSpin Express 4 Centrifuge: For in vitro diagnostic use to produce the rapid separation of whole blood contained in original collection tubes.

Reason for Recall

Iris International is recalling the RTH8 Rotor used in the StatSpin Express 4 Horizontal Centrifuge because the RTH8 rotor may develop cracks with use over time.

Recalling Firm

Iris Diagnostics — Chatsworth, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

3912 units

Distribution

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Norway, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Venezuela.

Code Information

Serial No. 3100 through 7012

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.

Back to All Device Recalls