Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of t...
FDA Device Recall #Z-1701-2014 — Class II — May 13, 2014
Recall Summary
| Recall Number | Z-1701-2014 |
| Classification | Class II — Moderate risk |
| Date Initiated | May 13, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Hologic, Inc. |
| Location | Danbury, CT |
| Product Type | Devices |
| Quantity | 3 units |
Product Description
Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.
Reason for Recall
Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
Distribution Pattern
Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
Lot / Code Information
Serial Numbers: 1800314 7263, 180031472, 18002023051PR
Other Recalls from Hologic, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1586-2026 | Class II | Brand Name: Aptima HPV Assay Product Name: Apt... | Feb 18, 2026 |
| Z-1095-2025 | Class II | The Rapid fFN Specimen Collection Tube (PN: 715... | Jan 8, 2025 |
| Z-0907-2025 | Class II | Panther Fusion GBS Assay REF PRD-04484 Kit cont... | Dec 13, 2024 |
| Z-0694-2025 | Class II | Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay, RE... | Nov 8, 2024 |
| Z-0782-2024 | Class II | PANTHER FUSION System REF 9121010000 The P... | Dec 15, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.