TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is ...

FDA Device Recall #Z-0993-2015 — Class II — August 21, 2014

Recall Summary

Recall Number Z-0993-2015
Classification Class II — Moderate risk
Date Initiated August 21, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maquet Medical Systems USA
Location Wayne, NJ
Product Type Devices
Quantity 6 units

Product Description

TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.

Reason for Recall

The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.

Distribution Pattern

US Nationwide Distribution including NE, NY and PA.

Lot / Code Information

part number: 1703.35AO, with software version up to 2.1.x.x.

Other Recalls from Maquet Medical Systems USA

Recall # Classification Product Date
Z-1831-2024 Class II CARDIOHELP-i System. Blood oxygenation and carb... Apr 15, 2024
Z-1077-2024 Class II BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Produc... Jan 16, 2024
Z-1076-2024 Class II Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5... Jan 16, 2024
Z-0977-2024 Class II BEQ-RF-32-USA RotaFlow Centrifugal Pump with BI... Jan 8, 2024
Z-0978-2024 Class II BO-RF-32-USA RotaFlow Centrifugal Pump with SOF... Jan 8, 2024

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.