PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 33CM, P/N 0250080694 Laparoscopic Manual Instruments are in...
FDA Device Recall #Z-0724-2015 — Class II — November 17, 2014
Recall Summary
| Recall Number | Z-0724-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 17, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Stryker Endoscopy |
| Location | San Jose, CA |
| Product Type | Devices |
| Quantity | 180,573 units total |
Product Description
PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 33CM, P/N 0250080694 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Reason for Recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot / Code Information
44493 647221 647992 745498 842129 845882 846787 847844 940092 940153 941265 941741 944824 1040124 1040628 1041467 1042526 1042527 1043031 1043779 10441058 1044108 1045039 1045371 1046186 1046395 1046955 1047068 1140260 1141156 1141790 1142614 1144485 1145002 1145148 1145969 1146672 1240949 1241420 1241798 1242459 1243202 1244181 1244438 1244527 1244594 1245525 1245545 1246645 1340475 1341523 1342388 1343130 1343554 1344027 1344278 1344686 1346102 1440693 1441635 1443223
Other Recalls from Stryker Endoscopy
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1566-2016 | Class II | PneumoSure XL High Flow Insufflator and Pneumo... | Mar 28, 2016 |
| Z-1831-2015 | Class II | SERFAS 90 degree Energy Probe, Part Number 279-... | Jun 3, 2015 |
| Z-0631-2015 | Class II | PKG, PENNINGTON FORCEPS, P/N 0250080242. Lapa... | Nov 17, 2014 |
| Z-0646-2015 | Class II | PKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250... | Nov 17, 2014 |
| Z-0679-2015 | Class II | PKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparo... | Nov 17, 2014 |
Frequently Asked Questions
Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.