Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TF...
FDA Device Recall #Z-0375-2016 — Class II — November 2, 2015
Recall Summary
| Recall Number | Z-0375-2016 |
| Classification | Class II — Moderate risk |
| Date Initiated | November 2, 2015 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Synthes (USA) Products LLC |
| Location | West Chester, PA |
| Product Type | Devices |
| Quantity | 5 |
Product Description
Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures. Intramedullary fixation rod, orthopedic.
Reason for Recall
TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.
Distribution Pattern
US Distribution to states of: CA, FL, MI and WI.
Lot / Code Information
Part Number Lot Number 04.037.162S 7950511
Other Recalls from Synthes (USA) Products LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0228-2026 | Class II | LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM... | Sep 26, 2025 |
| Z-0229-2026 | Class II | LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM... | Sep 26, 2025 |
| Z-1916-2024 | Class II | STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/... | Apr 25, 2024 |
| Z-1239-2024 | Class II | TFNA Femoral Nail 9mm, right, 130¿, L 235mm-In... | Jan 24, 2024 |
| Z-1238-2024 | Class II | RAPIDSORB IPS Battery Pack, Sterile-Intended fo... | Jan 24, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.