ARCHITECT Intact PTH Calibrators consists of 6 bottles (4.0 mL each). Calibrator A contains Bis T...

FDA Device Recall #Z-1671-2014 — Class II — February 12, 2014

Recall Summary

Recall Number Z-1671-2014
Classification Class II — Moderate risk
Date Initiated February 12, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Abbott Laboratories
Location Abbott Park, IL
Product Type Devices
Quantity 7,780 calibrator kits

Product Description

ARCHITECT Intact PTH Calibrators consists of 6 bottles (4.0 mL each). Calibrator A contains Bis Tris Propane buffer with protein (bovine) stabilizer. Calibrators B-F contain PTH (synthetic peptide) in Bis Tris Propane buffer with protein (bovine) stabilizer. Preservatives: sodium azide and ProClin 300. Product Usage: The ARCHITECT Intact PTH assay is an in vitro chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of intact parathyroid hormone (PTH) in human serum and plasma on the ARCHITECT i System.

Reason for Recall

Abbott has confirmed that a performance shift in the ARCHITECT Intact PTH assay has the potential to generate falsely elevated results on patient samples. Results generated with impacted lots may demonstrate a positive shift relative to those generated with previous calibrator and/or control lots. The issue may also impact your established ARCHITECT Intact PTH reference ranges. Abbott ARCHITECT In

Distribution Pattern

Worldwide Distribution - US Nationwide including District of Columbia and Puerto Rico; Algeria, Andorra, Angola, Argentina, Armenia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, ITALY, Jamaica, Japan, Jordan, Kenya, Korea (South Korea), Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mauritania, Mexico, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Panam¿, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, San Marino, Saudi, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, SPAIN, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Ukraine, Uruguay, Venezuela, Vietnam, and Yemen

Lot / Code Information

List Number: 8K25-01. Lot Numbers and Expiration Dates: Lot 00613I000, Expires 05 SEP 2014; Lot 01913C000, Expires 14 MAR 2014; Lot 01913K000, Expires 12 NOV 2014; Lot 02413E000, Expires 09 MAY 2014; Lot 01312K000, Expires 12 NOV 2013; Lot 00813A000, Expires 08 JAN 2014; Lot 01412H000, Expires 16 AUG 2013; Lot 00912F000, Expires 30 MAY 2013.

Other Recalls from Abbott Laboratories

Recall # Classification Product Date
Z-0961-2026 Class II Alinity ci-series System Control Module (SCM), ... Nov 13, 2025
Z-0489-2026 Class II Brand Name: Alinity hq Analyzer Product Name: ... Sep 18, 2025
Z-1652-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025
Z-1655-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025
Z-1653-2025 Class II Brand Name: CELL-DYN Ruby Product Name: CELL-D... Mar 18, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.