Maquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the pati...

FDA Device Recall #Z-1730-2014 — Class II — April 16, 2014

Recall Summary

Recall Number Z-1730-2014
Classification Class II — Moderate risk
Date Initiated April 16, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Maquet Medical Systems USA
Location Wayne, NJ
Product Type Devices
Quantity United States 1169 units

Product Description

Maquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the patient.

Reason for Recall

Maquet has identified an issue with the Maquet PowerLED 700 (certain serial numbers). One or several plastic fixation tabs can break, which can cause the ambient light module to partially detach.

Distribution Pattern

US Nationwide Distribution.

Lot / Code Information

Part number (begins with ALM) and associated serial numbers: ALM568300999 13, 14, 16, 19, 20, 21, 22, 23, 24, 25, 29, 30, 31, 32, 34, 35, 36, 41, 50 ALM568301999 13, 14, 16, 17, 20, 23, 24, 25, 26, 52, 54, 55, 56, 57, 60, 63, 64, 65, 66, 70, 71, 72, 74 ALM568302999 11, 43, 44, 46, 49, 50, 51, 54, 57, 58, 59, 6, 61, 63, 64, 66, 67, 71, 72, 74, 77, 78, 80, 81 ALM568303999 8, 9, 32, 52, 53, 58, 60, 61, 62, 65, 66, 69, 70, 71, 73, 74, 75, 77, 78, 79, 80, 81, 82, 85, 86, 87, 89, 91, 92, 94, 95, 96, 97, 99, 100, 105, 106, 108, 111, 118, 120, 121, 122, 159 ALM568320901 12012, 40015, 40017, 40018, 40019, 40020, 40021, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 50022, 50023, 50024, 50025, 50026, 50027, 50028, 50030, 50031, 50032, 50033, 50034, 50035, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50047, 50048, 50050, 50051, 50052, 50053, 50054, 50055, 50056, 50058, 50059, 50060, 50061, 50064, 50066, 50067, 50068, 50069, 50070, 50071, 50072, 70033, 70034, 70035 ALM568320905 40003, 40004, 40005, 40006, 40007, 40008, 40009, 40010, 40011, 40012, 40013, 40014, 40015, 40016, 50003, 50004, 50005, 50006, 50008, 50009, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50018, 50019, 50020, 50021, 60003, 60004, 70001, 70002, 70003, 70004, 70005, 70006, 70007, 70009, 70010, 70011, 70012, 70013, 70014, 70015, 70016, 70017, 70018, 70019, 70020, 70021, 70022, 70024, 70025, 70027, 70028, 70029, 70030, 70031, 70032, 70033, 70034, 70035, 70036, 70037, 70038, 70039, 70040, 70041, 70042, 70043, 70044, 70045, 70046, 70047, 70048, 70049, 70050, 70051, 70052, 70053, 70054, 70055, 70056, 70057, 70058, 70059, 70060, 70061, 70063, 70064, 70066, 70068, 70069 ALM568370900 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20039, 20040, 20041, 20042, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20054, 30038, 30039, 30053, 30054, 40006, 40010, 40012, 40013, 40014, 40015, 40016, 40018, 40019, 40020, 40027, 40028, 40029, 40030, 40031, 40032, 40033, 40034, 50000, 50001, 50004, 50006, 60007, 60008, 60009, 60010, 60012, 60031, 60041, 60042, 60043, 60044, 60046, 60047, 60048 ALM568370901 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20038, 20039, 20040, 20041, 20042, 20045, 20053, 20054, 20055, 20056, 20057, 20058, 20059, 20060, 20061, 20062, 20064, 20065, 20066, 20067, 20068, 20069, 20070, 32042, 40028, 40033, 40034, 40035, 40036, 40037, 40038, 40045, 50002, 50006, 50008, 50009, 50024, 50035, 50036, 50037, 50039, 50040, 50041, 50042, 50043, 50044, 50045, 50046, 50047, 50048, 60030 ALM568370902 20011, 20015, 20016, 20017, 20018, 20019, 20020, 20021, 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20031, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20040, 20041, 20042, 20048, 20049, 20050, 20051, 20057, 20058, 20059, 20060, 20061, 20062, 20063, 20064, 20067, 20068, 20069, 20070, 20071, 20072, 20073, 20074, 20075, 20076, 20077, 20078, 20079, 20080, 20083, 20086, 20087, 20089, 20090, 20091, 20094, 20095, 20096, 20098, 20104, 20105, 20106, 20107, 20108, 20109, 20110, 20111, 20112, 20113, 20114, 20115, 20116, 20117, 20118, 20119, 20120, 20121, 20122, 20123, 20124, 20125, 20127, 20132, 20133, 20135, 20136, 20137, 20139, 20140, 20141, 20144, 20151, 20152, 30000, 30051, 30053, 30054, 30062, 30063, 30065, 30066, 30067, 30068, 30078, 30081, 30090, 30091, 30092, 30099, 30107, 30127, 30128, 30130, 30131, 30134, 32164, 40000, 40006, 40007, 40009, 40011, 40012, 40013, 40014, 40015, 40017, 40018, 40019, 40020, 40021, 40023, 40039, 40040, 40041, 40042, 40043, 40044, 40045, 40046, 40047, 40048, 40050, 40051, 40052, 40053, 40054, 40055, 40056, 40057, 40058, 40060, 40064, 40065, 40067, 40078, 40081, 40082, 40083, 40084, 60030, 60031, 60036 ALM568370903 20022, 20023, 20024, 20025, 20026, 20027, 20028, 20029, 20030, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 20039, 20041, 20043, 20044, 20045, 20046, 20047, 20048, 20049, 20050, 20051, 20052, 20053, 20055, 20056, 20057, 20058, 20059, 20066, 20067, 20072, 20073, 20074, 20075, 20076, 20077, 20079, 20080, 20081, 20082, 20084, 20087, 20088, 20089, 20090, 20091, 20092, 20093, 20097, 20098, 20099, 30072, 40025, 50000, 50001, 50002, 50003, 50004, 50006, 50028, 50051, 50052, 50054, 50055, 50057, 50058, 50059, 50060, 60024, 60025, 60026, 60028 ALM568370904 20011, 20013, 20014, 20015, 20016, 20017, 20018, 20019, 20020, 20023, 20024, 20025, 20026, 20027, 20032, 20033, 20034, 20035, 20036, 20037, 20038, 30000, 30001, 30002, 30003, 30004, 30012, 30013, 30014, 30015, 30018, 30019, 30020, 30025, 30026, 30027, 30028, 30029, 30030, 30031, 30032, 30033, 30034, 30035, 30036, 30037, 30038, 30039, 30040, 30041, 30042, 30043, 40011, 40012, 40013, 40014, 40015, 40016, 40017, 40022, 40023, 40024, 40031, 40032, 40033, 40034, 40035, 40036, 50004, 50010, 50011, 50012, 50013, 50014, 50015, 50016, 50017, 50019 ALM568370905 20010, 20016, 20017, 20018, 20019, 20022, 20023, 30027, 40012, 40022, 40023, 40024, 40025, 40027, 40028, 40029, 40031, 40032, 40033, 40037, 40040, 40041, 40047, 40049, 40050, 50005, 50006, 50012, 50015, 50018, 50021, 60007, 60023, 60027, 60030, 60032, 60033

Other Recalls from Maquet Medical Systems USA

Recall # Classification Product Date
Z-1831-2024 Class II CARDIOHELP-i System. Blood oxygenation and carb... Apr 15, 2024
Z-1077-2024 Class II BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Produc... Jan 16, 2024
Z-1076-2024 Class II Model No. BEQ-HLS 5050 USA, HLS Set Advanced 5... Jan 16, 2024
Z-0977-2024 Class II BEQ-RF-32-USA RotaFlow Centrifugal Pump with BI... Jan 8, 2024
Z-0978-2024 Class II BO-RF-32-USA RotaFlow Centrifugal Pump with SOF... Jan 8, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.