Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the d...
FDA Recall #Z-0834-2015 — Class II — November 21, 2014
Product Description
Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
Reason for Recall
The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.
Recalling Firm
Intuitive Surgical, Inc. — Sunnyvale, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
1532 total both devices
Distribution
Nationwide Distribution.
Code Information
PN 950156-01
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated