Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is i...

FDA Device Recall #Z-1740-2013 — Class II — July 1, 2013

Recall Summary

Recall Number Z-1740-2013
Classification Class II — Moderate risk
Date Initiated July 1, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Stryker Instruments Div. of Stryker Corporation
Location Portage, MI
Product Type Devices

Product Description

Stryker TPS Universal Driver REF 5100- 99 Rx , The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Reason for Recall

The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Distribution Pattern

Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.

Lot / Code Information

Serial Number 0214200753

Other Recalls from Stryker Instruments Div. of Stryker C...

Recall # Classification Product Date
Z-1452-2022 Class II SurgiCount+ System Application, Software Versio... Jun 15, 2022
Z-0431-2021 Class II Stryker Zyphr Disposable Cranial Perforator, La... Oct 16, 2020
Z-1807-2020 Class II Neptune E-SEP 165mm Blade Electrode, Catalog Nu... Mar 26, 2020
Z-1402-2022 Class II Triton Canister System (finished part numbers F... Jan 22, 2020
Z-1097-2020 Class II SafeAir Telescopic Smoke Evacuation Pencil, Pr... Jan 13, 2020

Frequently Asked Questions

A software recall means the device's embedded software or firmware has a defect that could affect its performance or safety. Many software recalls are corrected through firmware updates that can be applied without physically replacing the device. For implantable devices, the update may be delivered wirelessly during a routine clinic visit. For external devices, the manufacturer may provide updated software files or replacement units. Contact your healthcare provider to determine whether your specific device and software version are affected and what action is recommended.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.