Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample,...
FDA Device Recall #Z-0890-2013 — Class II — February 12, 2013
Recall Summary
| Recall Number | Z-0890-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | February 12, 2013 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Oculus Innovative Sciences Inc |
| Location | Petaluma, CA |
| Product Type | Devices |
| Quantity | 56,566 1.5 oz tubes |
Product Description
Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for Sale. Pre-printed tube: MS31071, Shelf carton: PL51091-08 Atrapro Antipruritic HydroGel is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including radiation dermatitis and atopic dermatitis.
Reason for Recall
Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.
Distribution Pattern
Nationwide Distribution - One distributor in PA
Lot / Code Information
Lot 12A0012, Exp 2/2013; Lot 12A0013, Exp 2/2013: Lot 12A1116, Exp 2.2013; Lot 12A1119, Exp 2/2013; Lot 12B1144, Exp 3/2013; Lot 12B1148, Exp 3/2013; Lot 12C1157, Exp 4/2013; Lot 12E190. Exp 6/2013; Lot 12E194, Exp 6/2013; Lot 12E198, Exp 6/2013; Lot 12E202. Exp 6/2013; Lot 12E208, Exp 6/2013; Lot 12E211, Exp 6/2013.
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.