5.5mm x 40mm Noncannulated Screw Shank Subsequent Codes: KWP, KWQ, MNH, MNI
FDA Recall #Z-0370-2013 — Class II — January 5, 2009
Product Description
5.5mm x 40mm Noncannulated Screw Shank Subsequent Codes: KWP, KWQ, MNH, MNI
Reason for Recall
The tulip portion of the screw assembly can disengage from the shank after implantation.
Recalling Firm
NuVasive Inc — San Diego, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
240 units total
Distribution
Nationwide Distribution-including the states of OR, MO, FL, WI, CA, and NC.
Code Information
Batch: JK2037, JK2130, JK2037, JK2366
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated