NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. ...
FDA Recall #Z-0687-2013 — Class II — November 26, 2012
Product Description
NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. This tandem mass spectrometry (MSMS) reagent kit is intended for the measurement and evaluation of amino acids, succinylacetone, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. Quantitative analysis of these analytes ) and their relationship with each other is intended to provide analyte concentration profiles that may aid in screening newborns for metabolic disorders.
Reason for Recall
Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
Recalling Firm
Perkinelmer — Waltham, MA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
819 kits (10 plates/kit)
Distribution
Nationwide Distribution.
Code Information
Lot Numbers: 617253 , 617346, 618494 , 618653, 618764 , 619453
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.