KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highw...

FDA Device Recall #Z-2173-2013 — Class II — July 25, 2013

Recall Summary

Recall Number Z-2173-2013
Classification Class II — Moderate risk
Date Initiated July 25, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Roche Molecular Systems, Inc.
Location Branchburg, NJ
Product Type Devices
Quantity US-IVD: 350 Kits; CE-IVD: 441 KITS

Product Description

KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 South Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 USA The cobas¿ HPV Test is a qualitative in vitro test for the detection of Human Papillomavirus in patient specimens. The test utilizes amplification of target DNA by the Polymerase Chain Reaction (PCR) and nucleic acid hybridization for the detection of 14 high-risk (HR) HPV types in a single analysis. The test specifically identifies types HPV16 and HPV18 while concurrently detecting the rest of the high risk types (31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66 and 68). The cobas¿ HPV Test is indicated: (a) To screen patients 21 years and older with ASC-US (Atypical squamous cells of undetermined significance) cervical cytology test results to determine the need for referral to colposcopy. (b) To be used in patients 21 years and older with ASC-US cervical cytology results, to assess the presence or absence of high-risk HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. The results of this test are not intended to prevent women from proceeding to colposcopy. (c) In women 30 years and older, the cobas¿ HPV Test can be used with cervical cytology to adjunctively screen to assess the presence or absence of high risk HPV types. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. (d) In women 30 years and older, the cobas¿ HPV Test can be used to assess the presence or absence of HPV genotypes 16 and 18. This information, together with the physicians assessment of cytology history, other risk factors, and professional guidelines, may be used to guide patient management. Cervical specimens that may be tested with the cobas¿ HPV Test include the following liquid based collection media and collection device: " ThinPrep¿ Pap TestTM PreservCyt¿ Solution " Endocervical Brush/Spatula

Reason for Recall

During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas¿ 4800 HPV Master Mix reagent that was packaged in four (4) cobas¿ 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina

Distribution Pattern

Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zealand, Netherlands, Peru, Norway, Singapore, Romania, South Africa, Spain, South Korea, Sweden, Switzerland and UK.

Lot / Code Information

US-IVD: Material number 05232880190 Lot Number R13814, R15785 CE-IVD: Material Number: 05235898190 05235898190 05235901190 05235910190 Lot Number R15771, R15304

Other Recalls from Roche Molecular Systems, Inc.

Recall # Classification Product Date
Z-1104-2023 Class II cobas SARS-CoV-2 & Influenza A/B Qualitative as... Dec 21, 2022
Z-1106-2023 Class II cobas Influenza A/B & RSV UC (Utility Channel) ... Dec 21, 2022
Z-1105-2023 Class II cobas SARS-CoV-2 & Influenza A/B Qualitative nu... Dec 21, 2022
Z-0097-2023 Class II cobas 5800 instrument, Material No. 08707464001... Sep 8, 2022
Z-0828-2022 Class II cobas LiatSystem, respiratory virus panel nucle... Nov 16, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.