Hill-Rom 1000 Bed, Product No. P1160. The Hill-Rom 1000 Bed is intended for low to moderate acu...
FDA Device Recall #Z-1466-2013 — Class II — April 15, 2013
Recall Summary
| Recall Number | Z-1466-2013 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 15, 2013 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Hill-Rom, Inc. |
| Location | Batesville, IN |
| Product Type | Devices |
| Quantity | 8,737 devices |
Product Description
Hill-Rom 1000 Bed, Product No. P1160. The Hill-Rom 1000 Bed is intended for low to moderate acuity patient in the medical surgical area of the hospital.
Reason for Recall
Complaints have been reported for brakes disengaging when the bed is subjected to a hard jolt or force to the frame.
Distribution Pattern
Worldwide distribution: US (Nationwide) and countries of: Argentina , Bahamas, Bahrain, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, Ethiopia, France, Ghana, Guam, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, N. Mariana Islands, Pakistan, panama, Paraguay, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, United States, US Virgin Islands, Venezuela and Vietnam.
Lot / Code Information
Product number (P1160) with Serial Numbers J143AP0937 thru O080AP5661.
Other Recalls from Hill-Rom, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0783-2022 | Class II | (1)Traverse Rail Carrier E-System, wide 31017XX... | Jan 14, 2022 |
| Z-0639-2022 | Class II | Liko Mobile Lifts: Primarily intended for use i... | Dec 23, 2021 |
| Z-0640-2022 | Class II | Liko Mobile Lifts: Primarily intended for use i... | Dec 23, 2021 |
| Z-2600-2021 | Class II | LikoStretch 1900 Model Number: 3156051 - The co... | Aug 11, 2021 |
| Z-2599-2021 | Class II | Stretch Leveller Model Number: 3156200 - The co... | Aug 11, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.