Optilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is i...
FDA Device Recall #Z-2486-2012 — Class II — August 23, 2012
Recall Summary
| Recall Number | Z-2486-2012 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 23, 2012 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Biomet, Inc. |
| Location | Warsaw, IN |
| Product Type | Devices |
| Quantity | 363 units |
Product Description
Optilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.
Reason for Recall
Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw
Distribution Pattern
Worldwide Distribution-USA (nationwide) including the states of AL, GA, MT, TA, AZ, WA, CA, TA, IN, NY, NJ, VA, MA, MI, MD, PA, KY, NC, UT, NH, FL, OK, CO, LA, OH, MO and SD, and the countries of Japan, Netherlands, Australia, and Costa Rica.
Lot / Code Information
Catalog number: 36505; and lot numbers: 069540, 179190, 201610, 385380, 419940, 427530, 440360, 440370, 555370, 638560, 645820, and 956960.
Other Recalls from Biomet, Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0212-2026 | Class II | BIOMET ZipTight, Acute AC Joint Implant, Single... | Sep 23, 2025 |
| Z-2619-2025 | Class II | Juggerknot Mini Soft Anchors, Model/Catalog Num... | Sep 9, 2025 |
| Z-0477-2025 | Class II | JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model nu... | Oct 21, 2024 |
| Z-2179-2024 | Class II | RingLoc Bi-Polar Hip System Acetabular Cup, 41 ... | May 17, 2024 |
| Z-1607-2024 | Class II | Vanguard Knee System AS Tibial Bearing / DCM Ar... | Mar 13, 2024 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.