PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary t...
FDA Recall #Z-1077-2014 — Class II — December 9, 2013
Product Description
PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.
Reason for Recall
Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.
Recalling Firm
DePuy Orthopaedics, Inc. — Warsaw, IN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
2
Distribution
Worldwide Distribution-USA including the states of IA, CA, KY, and FL, and the countries of Finland, Germany, Sweden, and Czech Republic.
Code Information
Catalog Number 960015 Lot Numbers 7810268 and 7806929 (US) Lot Number 7806934 (OUS)
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.