Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-...
FDA Recall #Z-2106-2025 — Class II — June 10, 2025
Product Description
Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.
Reason for Recall
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Recalling Firm
Boston Scientific Corporation — Saint Paul, MN
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
14,724
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
Code Information
TF85-32-63-45 UDI-DI 00685447005391 Lots 35508747 35644132 35746496 35913449; TF85-32-63-55 UDI-DI 00685447005407 Lots 35508748 35738145
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.