NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT s...

FDA Device Recall #Z-1264-2019 — Class II — March 20, 2019

Recall Summary

Recall Number Z-1264-2019
Classification Class II — Moderate risk
Date Initiated March 20, 2019
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm GE Healthcare, LLC
Location Waukesha, WI
Product Type Devices
Quantity 16 units

Product Description

NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.

Reason for Recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Distribution Pattern

US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Lot / Code Information

Mfg. Lot or Serial # System ID 87CY63017 SHIPPED (AO 1630621) 87CX63007 AH0101NM05 87CX63014 030078NU28 87CX63008 190020NU01 87CX63006 030240NU11 87CX63015 705743NM870 87CX63011 M4184879 87CX63003 M9886873 87CX63004 X171582003 87CX63009 M171582018 87CX63010 UC1796NU01 87CX63013 NO1068MN18 87CX63005 A004NU20 87CX63002 C028NU01 87CX63012 C028NU02 87CX63001 900010NU12

Other Recalls from GE Healthcare, LLC

Recall # Classification Product Date
Z-1566-2022 Class II (1) GE Centricity Universal Viewer 6.0, Model 2... Jul 15, 2022
Z-1430-2022 Class II SIGNA Premier magnetic resonance scanner, model... Jun 24, 2022
Z-1466-2022 Class II Centricity PACS Software Version 7.0 SP0.0.4.7 Jun 14, 2022
Z-1333-2022 Class II GE Centricity Universal Viewer Zero Footprint. ... Jun 8, 2022
Z-1511-2022 Class II Service manuals for the CT, PET, or NM table as... Jun 2, 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.