Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL ...

FDA Device Recall #Z-0530-2015 — Class II — November 6, 2014

Recall Summary

Recall Number Z-0530-2015
Classification Class II — Moderate risk
Date Initiated November 6, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Abbott Medical Optics Inc (AMO)
Location Santa Ana, CA
Product Type Devices
Quantity 484,881 Nationwide (U.S. and Puerto Rico) and 483,984 Internationally

Product Description

Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL Sterile A Solution Sterile EO Packaging and Cannula Product of USA Abbott Healon Duet Dual Pack 0.55 ml OVD 0.85 ml Healon EndoCoat OVD Contents: -one sterile syringe of Healon Ophthalmic Viscosurgical Device (OVD) 0.55 ml (10 mg/ml Sodium Hyaluronate) -one sterile syringe of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) 0.85 ml (30 mg/ml Sodium Hyaluronate) Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 and 57550500 (packaged in Healon duet 10220010, 10220011 and 10220012)

Reason for Recall

AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result in the surgeon's hand unintentionally moving the cannula further than intended, resulting in potential trauma to the eye.

Distribution Pattern

Worldwide Distribution - USA (Nationwide and Puerto Rico) and internationally to the following countries; Italy, Austria, Saudi Arabia, Turkey, Hong Kong, Canada, Germany, Australia, Columbia, Venezuela, Netherlands, Portugal, New Zealand, Denmark, Argentina, France, Belgium, Costa Rica, Singapore, Ukraine, Guadeloupe, Kuwait, Chile, Peru, Mexico, Czech Republic, Serbia, Hungary, Slovenia, Guatemala, Finland, Honduras, Malaysia, United Kingdom, Spain, Switzerland, Israel, and Croatia.

Lot / Code Information

Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 57550500 (packaged in Healon Duet 10220010, 10220011, and 10220012) Healon EndoCoat Device identification: For Healon EndoCoat, each lot is assigned a unique lot number for traceability. The lot number is a 6-digit numerical code that is equal to the manufacturing order number assigned. The numbers 1-6 are sequentially increased representing the manufacturing order number. Example: Lot 024992. For Healon EndoCoat packaged in Healon Duet, each Healon Duet final product is identified witha unique batch number. The batch number consists of two alphabetic characters followed by five numeric characters. The first alphabetic character of each batch number designates the manufacturing site. The second alphabetic character of each batch number shall be sequential numbers starting with 30001. Example: Lot UP30729 (This lot number would represent the 729th unit manufactured by AMO in 2014).

Other Recalls from Abbott Medical Optics Inc (AMO)

Recall # Classification Product Date
Z-0929-2013 Class II AMO Sensar Intraocular Lenses (IOLs), Model AR4... Dec 20, 2012

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.