B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antib...

FDA Recall #Z-0921-2014 — Class II — December 13, 2013

Recall #Z-0921-2014 Date: December 13, 2013 Classification: Class II Status: Terminated

Product Description

B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

Reason for Recall

Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.

Recalling Firm

B. Braun Medical, Inc. — Allentown, PA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

30, 240 units

Distribution

US Distribution one importer/wholesale distributor in MO.

Code Information

Catalog #CT-2000-100lots 3H0125F151 and 3H0225F151, expiration 7/2018.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated