ILLlCO¿ MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Sp...

FDA Device Recall #Z-0168-2013 — Class II — September 25, 2012

Recall Summary

Recall Number Z-0168-2013
Classification Class II — Moderate risk
Date Initiated September 25, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Alphatec Spine, Inc.
Location Carlsbad, CA
Product Type Devices
Quantity 209 screws

Product Description

ILLlCO¿ MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal Pedical Fixation Orthosis, Spondylolisthesis Spinal Fixation The ILLICO MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. When used for a minimally invasive posterior approach ILLlCO¿ MIS Instrumentation is used in conjunction with polyaxial screw components. The ILLICO MIS Posterior Fixation System is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the ILLICO MIS Posterior Fixation System are dependent in part on the configuration of the assembled device and the method of attachment to the spine.

Reason for Recall

The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.

Distribution Pattern

Nationwide Distribution including AZ, IL, CA, SC, TN, TX, GA, CT, NY, FL, OK, LA, MO, NV,and UT.

Lot / Code Information

Product Codes: KW, MNI, MNH. Part NumberlLot Number: ¿PART DESCRIPTION LOT 7384S-30 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 30mm 646501 73845-35 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 35mm 646502 73845-40 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 40mm 646503 73845-45 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 45mm 646504 73855-35 Ti CANNULATED POLYAXIAL SCREW 5.5mm X 35mm 646510 73865-30 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 30mm 646515 73865-35 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 35mm 646516 73865-40 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 40mm 646517 73865-45 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 45mm 646542 73865-50 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 50mm 646543 73875-35 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 35mm 646546 73875-45 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 45mm 646548 73875-50 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 50mm 646549

Other Recalls from Alphatec Spine, Inc.

Recall # Classification Product Date
Z-1342-2026 Class II ATEC Lateral Navigation Disc Prep Instruments ... Jan 2, 2026
Z-1343-2026 Class II ATEC Lateral Navigation Disc Prep Instruments ... Jan 2, 2026
Z-1065-2025 Class II Calibrate CCX Interbody System for spinal fusio... Dec 15, 2024
Z-1182-2024 Class II IdentiTi ALIF Standalone Interbody System: Par... Jan 26, 2024
Z-2475-2023 Class II LIF AMP, Adjustable Awl, REF 117-165, Part of t... Jul 7, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.