Chromic Gut Absorbable Sutures, SURGIGUT* 0 CHR GS-25 90CM X36 (CG-904), SURGIGUT* 1 CHR GS-24 90...

FDA Recall #Z-1388-2023 — Class II — March 1, 2023

Recall #Z-1388-2023 Date: March 1, 2023 Classification: Class II Status: Ongoing

Product Description

Chromic Gut Absorbable Sutures, SURGIGUT* 0 CHR GS-25 90CM X36 (CG-904), SURGIGUT* 1 CHR GS-24 90CM X36 (CG-915), CHROMIC GUT 4-0 75CM V20 (GG-121), CHROMIC GUT 3-0 75CM V20 X36 (GG-122), CHROMIC GUT 2-0 75CM V20 X36 (GG-123), CHROMIC GUT 2-0 75CM V26 (GG-127), CHROMIC GUT 4-0 45CM P12 X36 (SG-5637), CHROMIC GUT 5-0 45CM P-13 X12 (SG-5687G), CHROMIC GUT 5-0 75CM CV23 (UG-202), CHROMIC GUT 4-0 75CM CV23 X36 (UG-203)

Reason for Recall

Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.

Recalling Firm

Covidien, LP — North Haven, CT

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

36204 units

Distribution

Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.

Code Information

GTIN: 20884521054001, 20884521054056, 20884521054889, 20884521054896, 20884521054896, 20884521054902, 20884521054902, 10884521054905, 20884521054940, 20884521055688, 20884521055770, 20884521056135, 10884521056145, 20884521056142, 20884521056142, 20884521056142; Lot: D2H1593ZY D2H1022ZY D2F1106ZY D2H1607ZY D2H1609ZY D2H0819ZY D2H2053ZY D2H2053ZY D2H2055ZY D2H1435FZY D2H0829FZY D2H0830FZY D2H0381ZY D2H1523ZY D2H1584ZY D2H1584ZY D2H1587ZY D2H1589ZY

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated